IO0064 - VENTICELL 222 R

  • - Accurate and quick laboratory oven
  • - 222-liter chamber volume
  • - Working temperature of from +10°C over ambient temperature up to 250°C300 °C
  • - Patented system of forced air flow in the chamber ensures a homogenous temperature profile for all processes of drying, heating and sterilizing materials in laboratories
  • - High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
  • - Greater speed and accuracy of all tempering procedures ensures cost-saving operation; suitable especially for materials of high humidity, for demanding and accurate tests and processes
  • - Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
  • - Wide range of optional equipment and accessories
  • - Output for a printer or PC connection
  • - Validation possibility (IQ , OQ)

Technical dataVenticell_55_standard_1

Interior of stainless steel material No. 1.4301
  • volume 222 l
  • depth 540 mm
  • width 540 mm
  • height 760 mm

Tray

  • standard equipment 2 pcs
  • storage area (w × d) 520×485 mm
  • racks 10 max. number
  • distance between guide rails 70 mm

Maximal weight of the load *)

  • inside the oven 70 kg/oven
  • per tray 30 kg/tray

Number of doors

  • 1 pcs

External dimensions (including door and handle)

  • depth 790 mm
  • diameter of the air branch outer/inner 52/49 mm
  • width 760 mm
  • height, incl. Feet (N) and Caster (K) 1090 (N) mm

Package dimensions (three layers carton)

  • depth 860 mm
  • width 830 mm
  • depth 1280 mm

Weight

  • gross 116 kg
  • net 100 kg

Working temperature **) (beginning of the regulation) - VENTICELL / VENTICELL plus

  • from 10 °C above ambient temp. to 250 / 300 °C

Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door, > 50 °C

  • time variation cca (±) 0,4 °C
  • space deviation (±) % cca (±)% of the reached temperature 1 %

Time required to reach 250 °C with closed air flap and voltage 230 V

  • 70 min

Heat emission at 250 °C

  • 990 W

Air exchange speed at 250 °C

  • 24 cca/hour

Electricity - mains 50/60 Hz

  • max. power input 1,9 kW
  • nominal voltage 230 V
  • current 8,3 A
  • stand by 5 W 

*) Approx. 50 % of the tray area can be filled while maintaining uniform air circulation inside the chamber.
**) Standard type is up to 250 °C, optional type is up to 300 °C.
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.

Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters