II0035 - INCUCELL V 55 - Standard

  •  - Natural, precise – laboratory incubator
  • - 55-liter chamber volume
  • - Working temperature of from +10°C above ambient up to 99,9°C
  • - Patented system of air flow in the chamber ensures quicker and more accurate distribution of the preset temperature in the chamber, which enables ideal heating of the samples
  • - High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening; the standard equipment includes inner glass door
  • - Model with a fan ensures quicker and more accurate distribution of temperature with small deviations; suitable for demanding and accurate tests and samples cultivation
  • - Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
  • - Wide range of optional equipment and accessories
  • - Output for a printer or PC connection
  • - Validation possibility (IQ , OQ) 

Technical dataVenticell_55_standard_1

Interior of stainless steel material No. 1.4301
  • volume 55 l
  • depth 370 mm
  • width 400 mm
  • height 350 mm

Tray

  • standard equipment 2 pcs
  • storage area (w × d) 380×335 mm
  • racks 4 max. number
  • distance between guide rails 70 mm

Maximal weight of the load *)

  • inside the oven 50 kg/oven
  • per tray 20 kg/tray

Door

  • 1 pcs

External dimensions (including door and handle)

  • depth 640 mm
  • diameter of the air branch outer/inner 52/49 mm
  • width 620 mm
  • height, incl. Feet (N) and Caster (K) 680 (N)

Package dimensions (three layers carton)

  • depth 730 mm
  • width 700 mm
  • depth 880 mm

Weight

  • gross 66 kg
  • nett 55 kg

Working temperature (beginning of the regulation)

  • from 10 °C above ambient temp. to: 99,9 °C

Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door

  • time variation cca (±) 0,2 °C
  • space deviation cca (±) 0,3 °C

Time required to reach 37 °C with closed air flap and voltage 230 V

  • 41 min

Heat emission at 37 °C

  • 80 W

Air exchange speed at 37 °C

  • 45 per hour

Electricity - mains 50/60 Hz

  • max. power 0,7 kW
  • nominal voltage 230 V
  • current 3 A
  • power by stand by 5 W
*) Approx. 50 % of the tray area can be filled while allowing uniform air circulation inside the chamber.
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.