II0034 - INCUCELL V 22 - Standard

  • - Natural, precise – laboratory incubator
  • - 22-liter chamber volume
  • - Working temperature of from +10°C above ambient up to 99,9°C
  • - Patented system of air flow in the chamber ensures quicker and more accurate distribution of the preset temperature in the chamber, which enables ideal heating of the samples
  • - High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening; the standard equipment includes inner glass door
  • - Model with a fan ensures quicker and more accurate distribution of temperature with small deviations; suitable for demanding and accurate tests and samples cultivation
  • - Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
  • - Wide range of optional equipment and accessories
  • - Output for a printer or PC connection
  • - Validation possibility (IQ , OQ)

Technical dataMMM Incucell Incubator

Interior of stainless steel material No. 1.4301

  • volume 22 l
  • depth 320 mm
  • width 240 mm
  • height 295 mm

Tray

  • standard equipment 2 pcs
  • storage area (w × d) 185×265 mm
  • racks 4 max. number
  • distance between guide rails 60 mm

Maximal weight of the load *)

  • inside the oven 25 kg/oven
  • per tray 10 kg/tray

Door

  • 1 pcs

External dimensions (including door and handle)

  • depth 580 mm
  • diameter of the air branch outer/inner 52/49 mm
  • width 406 mm
  • height, incl. Feet (N) and Caster (K) 604 (N)

Package dimensions (three layers carton)

  • depth 665 mm
  • width 465 mm
  • depth 665 mm

Weight

  • gross 36 kg
  • nett 31 kg

Working temperature (beginning of the regulation)

  • from 10 °C above ambient temp. to: 99,9 °C

Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door

  • time variation cca (±) 0,1 °C
  • space deviation cca (±) 0,4 °C

Time required to reach 37 °C with closed air flap and voltage 230 V

  • 8 min

Heat emission at 37 °C

  • 70 W

Air exchange speed at 37 °C

  • 45 per hour

Electricity - mains 50/60 Hz

  • max. power 0,96 kW
  • nominal voltage 230 V
  • current 4,2 A
  • power by stand by 5 W