II0031 - INCUCELL 55 - Standard

  • - Natural, precise – laboratory incubator
  • - 55-liter chamber volume
  • - Working temperature of from +5°C above ambient up to 99,9°C
  • - Noiseless operation and fine air flow in the chamber
  • - High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
  • - High-quality and safe drying, heating and tempering of common laboratory materials in laboratories and industry; suitable for safe processing of microbiological cultures
  • - Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
  • - Wide range of optional equipment and accessories
  • - Output for a printer or PC connection
  • - Validation possibility (IQ , OQ)

Technical data

Interior of stainless steel material No. 1.4301       Incucell_55_standard_2

  • volume 55 l
  • depth 370 mm
  • width 400 mm
  • height 350 mm

Tray

  • standard equipment 2 pcs
  • storage area (w × d) 380×335 mm
  • racks 4 max. number
  • distance between guide rails 70 mm

Maximal weight of the load*)

  • inside the oven 50 kg/oven
  • per tray 20 kg/tray

Door

  • 1 pcs

External dimensions (including door and handle)

  • depth 640 mm
  • diameter of the air branch outer / inner 52/49 mm
  • width 620 mm
  • height, incl. Feet (N) and Caster (K) 680 (N)

Package dimensions (three layers carton)

  • depth 730 mm
  • width 700 mm
  • depth 880 mm

Weight

  • gross 66 kg
  • nett 55 kg

Working temperature (beginning of the regulation)

  • from 5 °C above ambient temp. to: 99,9 °C

Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door

  • time variation cca (±) 0,2 °C
  • space deviation cca (±) 0,5 °C

Time required to reach 37 °C with closed air flap and voltage 230 V

  • 49 min

Heat emission at 37 °C

  • 30 W

Air exchange speed at 37 °C

  • 5 per hour

Electricity - mains 50/60 Hz

  • max. power 0,3 kW
  • nominal voltage 230 V
  • current 1,3 A
  • power by stand by 5 W

*) Approx. 50 % of the tray area can be filled while allowing uniform air circulation inside the chamber.
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.