DISTILLATION UNIT, AUTOMATIC WITH TITRATOR CONNECTION, 230V 50-60Hz UDK149


ID8050 

ID8050 thumb

Automated Operation

  • Programmable boric acid, water, sodium hydroxide addition and distillation time
  • Automatic processing in combination with a variety of titrators
  • Selectable steam generation output level 10% – 100%
  • Auto removal of residues from titrator & sample tube

High Tech

  • 3.5 inch colour touch screen
  • Library of 20 customisable methods
  • Choice of 6 languages
  • Mouse & printer connections
  • No loss of nitrogen: distillate temperature stays below threshold
  • Accurate dosing of reagents with high precision pumps

Versatile Data Handling

  • Powerful archiving features
  • Data exporting via USB to pen drive or PC in .xls format
  • Data manipulation software can be user preferred

Outstanding Reliability

  • Maintenance-free, resistant, patented steam generator
  • Patented titanium condenser for unmatched performance
  • Durable technopolymer splash head
  • Chemically-resistant technopolymer housing

Operator Safety

  • Lever operation simplifies tube insertion / removal (all sizes)
  • Protection screen for distillation chamber
  • Drip tray collects splashes
  • Zero pressure steam generator

Ecology & Standards

  • Performs to AOAC, EPA, DIN and ISO standards
  • Incorporates TEMS™ technology
  • Automated water supply control reduces consumption

GENERAL FEATURES

Structure:Corrosion-resistant technopolymer
Display:3.5" colour touch screen
Time related to start for Devarda alloy analysis:0 – 99 min
Protocol library:20 customisable methods
Reproducibility (RSD):≤ 1 %
Recovery:≥ 99.5 % at nitrogen levels between 1 and 200 mg N
Detection Limit:≥ 0.1 mg N
Power:2100 W
Weight:27 kg / 59.5 lb
Dimensions (WxHxD):385x780x416 mm (15.2x30.7x16.4 in)

PERFORMANCE

Distillation time:3 minutes to collect 100 ml of distillate
Sodium hydroxide addition:Automatic
Water addition:Automatic
Boric acid addition:Automatic
Sodium hydroxide volume:0 - 150 ml
Water volume:0 - 200 ml
Boric acid volume:0 - 100 ml
Distillation time:Settable
Distillation residues removal:Automatic
Steam flow regulation:From 10 to 100 %
Tap water consumption:From 0,5 l/min at 15 °C – from 1 l/min at 30 °C
Connections:Ethernet, 2 x USB, RS232 and TTL for PC, printer, titrator, mouse, pen drive
On-board archive for data storage